Specification:
Voltage |
AC Voltage Range | 57.7V-660V |
Power of the Voltage output | MAX.30VA |
Resolution | 0.01% |
Accuracy | 0.1% |
Stability | 0.01%/120s |
Distortion factor | 0.2% |
Current |
Test current output | 1A-50A |
Power of the current output | Max. 100VA |
Resolution | 0.01% |
Accuracy | 0.05% |
Stability | 0.01%/180s |
Distortion factor | < 0.3% for linear resistance load |
Phase angle |
Range | 0 -- 360° |
Resolution | 0.01° |
Accuracy | 0.2° |
Frequency |
Range | 40Hz – 70Hz |
Resolution | 0.005Hz |
Pulse output |
Harmonic Output | 2-63 |
Pulse input |
Input channel | 3 |
Input frequency | Max. 20Hz |
Other |
Voltage Supply | 220V±15% /110V±15% (optional) |
Ambient temperature | 20°C -- 30°C |
Relative humidity | ≤85% |
Dimensions (W×H×D) (mm) | 500 X 440 X 180 |
Weight (kg) | 46(5C)/22(3C) |
Description:
YC99T series is a portable three-phase meter tester (calibrator), suitable for laboratory and local use. YC99T series calibrator utilises the testing technology with the lasted development of energy meter calibration. YC99T5C has the same software function as YC99T-3C calibrator, but different voltage and current output.
YC99T-5C calibrator supports IEC standard communication protocol, and can complete demand indication error, demand period error, time switching error, rate period energy indication error and other multi-function test projects. YC99T-5C calibrator has a constant temperature quartz oscillator, which can verify the daily time error of electric energy meter.
YC99T-5C is designed with fantastic panel screen and wiring arrangement. The front and rear panels are designed to be easy to operate.
YC99T-5C not only provides normal meter testing, it can also be used as harmonic power output equipment, which is able to generate stable harmonic output.
YC99T Series Portable Calibrator is capable of detecting daily time error of energy meter.
1. The error of voltage, current, active power is ±0.05%, the error of reactive power is ±0.1%.
2. The stability of voltage, current, active power is 0.01%RG/3min, the distortion percentage of alternative voltage and alternative current : ≤0.2%.
3. The voltage range of YC99T-5C from 57.7V-660V, current range:1A-50A
4. Harmonic output range:2-63, harmonic output is under measurement and analysis.
5. Test data can be managed by computer via RS232 interface
6. YC99T-5C can not only be operated by instrument, but also controlled by PC.
7. Calibrate energy meters automatically, including items such as, preheating, starting, creeping, error, influence, quantity test etc.
8. MUTs can be accessed via RS485, RS232 and Ethernet.
9. Diagnostic function of functional fault and internal module communication.