Add to Cart
The YC1893D meter test system is a full-automatic test system for testing and calibration of both single-phase and three-phase electric energy meters, which composes of:
The YC1893D meter test system is fully compliant with the IEC 60736 standard and is suitable for testing meters according to the following standards:
The test bench is multiple positions up to 3 meters, 0.02% of accuracy class Three-phase Meter Test System with ICTs, with multiple serial port communication system for parallel communication for testing of smart meters.
Voltage | 3×240±15% |
Frequency | 50 Hz ±15% |
Power capacity | > 6 kVA |
Power efficient(at full load) | ≥85% |
Ambient temperature | -10 °C -- +40 °C |
Relative humidity | 35% -- 85% |
Voltage | |
Test voltage output (Phase-Neutral) | 3×(24 V – 480 V) or 600 V F-F customized |
Power of the voltage output | 3×100VA |
Resolution | Better than 0.01% of full scale value of range |
Setting accuracy | Better than 0.02% at the final range value |
Stability | Better than 0.005%/h (integration time 150 s) |
Load regulation from 0-max. load | < 0.01% |
Distortion factor | < 0.3% for linear resistance load |
Harmonics | 2nd – 21st free programmable |
Current | |
Test current output | 0.3 mA – 120 A |
Power of the current output | 3×150VA |
Resolution | Better than 0.01% of full scale value of range |
Setting accuracy | Better than 0.02% at the final range value |
Stability | Better than 0.005%/h (integration time 150 s) |
Load regulation from 0-max. load | < 0.01% |
Distortion factor | < 0.3% for linear resistance load |
Harmonics | 2nd – 21st free programmable |
Phase angle | |
Range | 0 -- 360° |
Resolution | 0.01° |
Setting accuracy | 0.1° |
Frequency | |
Range | 45 Hz – 65 Hz |
Resolution | 0.01 Hz |